Biopsy Depth Measurement - GE LOGIQ A100 Technical Publication

Table of Contents

Advertisement

Basic Measurements

Biopsy Depth Measurement

To measure the desired depth for needle insertion during
a biopsy:
Press CONTROL B ENTER or the alphanumeric key 'B'
Ctrl
to turn the Biopsy Zone Display ON.
Enter
B
Ctrl
Enter
Press the MEASUREMENT key to invoke the biopsy
Meas
depth. A dotted line appears in the middle of the zone
and an open ended " " cursor appears on this line. The
position of the cursor gives the biopsy depth which is
displayed on the left side of the image display.
Use the TRACKBALL to move the cursor along the
center line. The biopsy depth values are computed and
TB
displayed on the left side of the image display.
Press CLEAR to exit out of the measurement mode.
Ctrl
Enter
Press CONTROL B ENTER or the alphanumeric key 'B'
to exit the Biopsy Zone Display.
B
122
t
α 100 User Manual
LOGIQ
2211157–100 Rev 0