Probe Cleaning Process (
.
Special Cleaning Instructions for the E72: When
cleaning the E72 probe, it is important to be sure that all
surfaces are thoroughly cleaned. This probe has an
adjustable two-part handle that must be disassembled to
gain access to all surfaces. To disassemble the handle,
completely remove the handle adjustment screw located
mid-way between the cable entry and probe tip the two
handle halves and adjustment screw must be thoroughly
cleaned along with the main probe shaft as described
earlier in step 2. After rinsing and drying is completed,
the probe handle can be loosely reassembled for the
disinfection process.
C36
C55
C31
Fluid Level
Illustration 54. Probe Immersion Levels
264
)
cont'd
3. Rinse the probe with enough clean potable water
to remove all visible soap residue.
4. Air dry or dry with a soft cloth.
L76
VE5
Contact face within
Aperature
Patient Environment
E72
t
α 100 User Manual
LOGIQ
2211157–100 Rev 0