Cardiac Measurements

Amplitude measurement

The default measurement in M
amplitude
+
The amplitude of a moving cardiac structure is the
+
greatest distance it has moved in a well defined phase of
the cardiac cycle and is expressed in millimeters.
Amplitude measurements are done only in M
Perform a scan in B/M or M
M
A "
Press FREEZE to stop image acquisition.
Meas
Press the MEASUREMENT key once to display the
measurement menu and an open ended " " cursor
appears on the screen.
Use the TRACKBALL to move the open ended " " cursor
to the amplitude measurement start point.
Press set to fix the measurement start point. A second
cursor appears. Amplitude in mm appears on the left
side of the screen.
Use the TRACKBALL to move the second open ended
TB
" " cursor to the measurement end point.
Press SET to complete the measurement. The amplitude
value is displayed in mm on the left of the screen.
Repeat the same procedure to make more amplitude
measurements, if required. Up to 4 measurements will
be displayed in mm at a time.
224
Mode is amplitude .
Mode.
Mode.
t
α 100 User Manual
LOGIQ
2211157–100 Rev 0