Preparing the System for Use
Power Cord (cont'd)
To connect the system to the power supply:
To avoid risk of fire, the system power must be supplied
WARNING
from a separate properly rated outlet.
The system is supplied with a power cord. Under no
circumstances should this cord be altered or changed.
To assure grounding reliability, connect a 'hospital grade'
or 'hospital only' grounded power outlet.
Plug
Outlet
115 VAC, 140 VA
Plug and Outlet Configuration
(USA)
Illustration 4. Example Plug and Outlet Configurations
18
1. Ensure that the wall outlet is of the appropriate
type.
2. Make sure that the power switch is turned off.
3. Unwrap the Power Cord. Ensure sufficient slack
in the cable so that the plug is not pulled out of
the wall if the system is moved slightly.
4. Push the power plug securely into the wall.
Plug and Outlet Configuration
220 VAC, 170VA
(Europe)
t
α 100 User Manual
LOGIQ
2211157–100 Rev 0