2.7 Self-check

2.7.1 Self-check Procedure

To run the self-check of the emulator, do so as explained here below. While the self-check is in progress, the LEDs will change
as shown in Figure 2.7.
(1) If the user system is connected, disconnect it.
(2) Set the switches in the emulation pod to the factory settings as shown in Figure 2.8.
(3) Within 2 seconds of activating power to the emulator, press the system reset switch on the emulator front panel to switch
the emulator to maintenance mode.
(4) Check the SAFE LED starts blinking and then press the system reset switch again.
(5) The self-check will start. If the normal result is displayed in about 30 seconds, the self-check has terminated normally.
POWER
SAFE
ERROR
S T A T US O F
SYSTEM
RESET
(1) All the LEDs light
Immediately after turning on the power
Figure 2.7 LED display in the self-check procedure
REJ10J0964-0100 Rev.1.00 August 01, 2005
POWER
POWER
CLOCK
SAFE
RESET
ERROR
RUN
S T A T U S O F
SYSTEM
HALT
RESET
: On
: Off
: Blinking
POWER
CLOCK
RESET
RUN
HALT
(2) SAFE LED blinks
Self-check
POWER
POWER
SAFE
CLOCK
ERROR
RESET
RUN
S T A T US O F
SYSTEM
HALT
RESET
Completed normally
POWER
POWER
SAFE
CLOCK
ERROR
RESET
RUN
S T A TU S O F
SYSTEM
HALT
ERROR LED lights
RESET
System status error
POWER
POWER
SAFE
CLOCK
ERROR
RESET
RUN
S T A TU S OF
SYSTEM
HAL T
SAFE and ERROR LED blink
RESET
Target status error
2. Setup
Page 29 of 88

Advertisement

Table of Contents
loading

Table of Contents