Sensor Settings; Adjustment - sauter FS Series Manual

Table of Contents

Advertisement

11.3 Sensor settings

Select the channel on which you want to make settings. Only external channels to
which a sensor is connected are displayed. Only use the plugs with memory chip fitted
by SAUTER for external sensors.
Force gauges and optional external sensors, as described in chapter 13.2 are
delivered adjusted. Only make adjustments on this page if your device no longer
displays the correct (and checked) measured values and you have experience with the
adjustment of force measuring devices. Calibrated measuring instruments must be
recalibrated after an adjustment.

11.3.1 Adjustment

To recalibrate a device, you need a calibration weight. Ideally with the nominal load of
the sensor to be adjusted.
11.3.1.1 Measuring range and unit
Do not change these values. A new adjustment and possibly also a recalibration will
be necessary!
11.3.1.2 Readability
The displayed value is the smallest adjustable readability. Set the readability required
for your application 11.2.2 chapter, set the readability required for your application
before using the meter.
11.3.1.3 Two-point adjustment
Click on Adjustment and then on Two-Point Adjustment.
11.3.1.3.1 Enter the values of the calibration weights
Enter your adjustment points in the upper part of the display at "Value? For example:
-
Positive adjustment point: 500N
-
Negative adjustment point: 0N
It is recommended to adjust with a weight that is close to the nominal value. For
example, a weight of 400N to 500N for a force gauge with a maximum measuring range
of 500N. Too small adjustment weights lead to a high measuring error!
22
FS-BA-en-2210.docx

Advertisement

Table of Contents
loading

Table of Contents