Version 1.4 12/2016 Instruction Manual Digital Force Gauge Thank you for buying a digital SAUTER force gauge. We hope you are pleased with your high quality force gauge with its big functional range. If you have any queries, wishes or helpful suggestions, do not hesitate to call our service number.
1 Scope of delivery Sensor Inside SAUTER FK Electrical power supply Transport case Standard attachments, as shown below: Extension rod: 90mm Sensor inside signifies that the measuring cell is inside the housing of the display unit Conditions: Measurement principles: 1. Continuous measurement (Track) 2.
2 Working Conditions Working temperature: 10°C to 30°C Humidity: 15% to 80% 3 Electrical power supply - Either by batteries (6 x AA batteries) not rechargeable! Batteries have to be changed if necessary. - or current power supply Using electricity network: Connection by power adapter cable 4 Technical data Accuracy: ±...
To activate please press: MAX key 6 Backside Fixing to a test stand Mounting is possible on all SAUTER test stands by means of 4 x M3 screws. 7 Warning notices Force measurements which are performed in an incorrect way, can lead to serious injuries of living beings and damage of objects.
Page 6
- The instrument may only be used in compliance with the described guidelines. - Varying ranges of application or planned use must be approved by SAUTER in writing. Warranty The warranty is not valid under following aspects: •...
For this purpose, the responsible user must define a suitable interval as well as the nature and scope of this inspection. Information is available on SAUTER home page (www.sauter.eu) concerning the monitoring of instrument test substances and the test weights required for this.
“CAL” appears on the display. 3. Wait a few seconds and display will show the necessary adjustment weight, e.g. 5,000kg for FK 50 or 10,00kg for FK 100. 4. Put the corresponding weight at the loading hook. Avoid trembling and keep still while holding the weight.