Adjusting Foils; General Notes; 10. Restore Factory Settings - sauter TC 1250-0.1F Instruction Manual

Digital coating thickness gauge
Table of Contents

Advertisement

a) When the battery symbol ``+/-`` appears on the display, the batteries should be
replaced.
b) The battery cover 4-8 is removed from the meter and the batteries are removed.
c) The batteries (4x1.5 V AAA/UM-4) are inserted by observing the polarity when
inserting them.
d) If the device is not used for a longer period of time, the batteries should be removed.

8. Adjusting foils

This instrument includes an adjustment foil set with different foils and measuring
ranges, but always covers the measuring range from 20 to 2000µm. These are also
available as optional accessories, article ATB-US07

9. General notes

a) The measuring instrument should always be adjusted on the base material used for
the actual measurement, if possible, instead of on the supplied zero plate F made of
steel or N made of aluminium. This makes the measuring accuracy more accurate from
the outset.
b) The measuring probe may wear out. The lifetime of it usually depends on the number
of measurements and the roughness of the layer to be measured. The replacement of
a measuring probe should only be carried out by qualified personnel
can be carried out.

10. Restore factory settings

a) It is recommended that the factory settings be performed in the following cases:
- no more measurements can be made at all
- the measuring accuracy is impaired due to drastic changes in the environment in
which the measuring instrument is used
- After replacing the measuring probe
(b) Procedure:
The factory settings include both the settings for "Fe" (F) and for "NFe" (N).
Please note which symbol is shown on the display.
Only one setting at a time can be made, as follows:
- Press the on/off button 4-6 and do not release it until the display shows "CAL". It
takes about 5 seconds from the start of this button operation until CAL is displayed.
The measuring probe 4-1 is carefully pressed onto the corresponding carrier material.
Then the zero key is pressed, which is accompanied by a beep.
When NF:H appears on the display, the measuring probe 4-1 is raised by more than 5
cm.
Then the zero button is pressed again and the meter returns to the measuring mode.
The factory settings are now complete.
Note: This process should always be completed within the following 6 seconds.
Otherwise, it will be automatically interrupted by the unit and the reset will be invalid.
TC-BA-e-2020
7