7.2 Connection sockets for external sensors

Depending on the model, 2 or 4 strain gauge sensors can be connected to your force
gauge and measured simultaneously. There are always 5 sockets on your device. The
measuring channels 3 and 4 and the interface socket are only optionally assigned!
Measuring channel
CH4 (optional)
Measuring channel
CH3 (optional)
You will receive the connectors when you order a sensor from SAUTER, as described
in section 13.2 "" described.

7.2.1 Compatibility and necessary accessories

It is recommended to purchase the sensors directly from SAUTER. There you will
receive compatible sensors with the corresponding necessary services. The services
listed in chapter 13.2 include the adjustment of the measuring chain, the connector and
the required data memory chip.
7.2.2 Plug & play from external sensors
External sensors can be unplugged and replaced by other adjusted sensors. Due to
the tolerances of the respective input channels, the sensors must be adjusted with the
device used. For this purpose, the unit must be sent to the KERN laboratory when
sensors are reordered.
If possible, always connect the sensor to the channel on which it was
adjusted and calibrated. It is best to keep a record of this! This will give you
the highest possible accuracy.