Congratulations on the purchase of a digital force measuring device with internal measuring cell from SAUTER. We hope you will enjoy your quality measuring device with a wide range of functions. Please do not hesitate to contact us if you have any questions, requests or suggestions.
If any damage is evident, please notify SAUTER GmbH immediately. SAUTER offers optional software and accessories to make the measuring instrument more versatile in use. Please ask SAUTER or the SAUTER supplier or visit our website www.sauter.eu. Scope of delivery...
Technical data 3.1 Technical data FK with internal loadcell up to 500N Measuring FK 10 FK 25 FK 50 FK 100 FK 250 FK 500 FK 1k device Capacity 100N 250N 500N 1000N Work 10°C to 30°C temperature Relative air...
Battery operation / power supply Optionally available in mains or battery operation Mains operation: Connection via mains adapter to mains power supply Battery operation for mobile use: 6x AAA batteries, included in delivery not rechargeable! Batteries must be replaced if necessary Display indication Position Description...
Inverting the display by 180° If the device itself is rotated by 180°, the display also rotates automatically. Warnings Incorrectly performed force measurements can lead to serious injury to persons and damage to objects and must therefore only be performed by trained and experienced personnel.
Page 7
- The measuring instrument may only be operated or maintained by trained personnel. - The measuring instrument may only be used in accordance with the described specifications. - SAUTER must give written approval for any other areas of use / applications. Warranty Warranty expires in case of...
Page 8
Information on the monitoring of measuring instruments and the necessary test weights is available on the SAUTER homepage (www.sauter.eu). The weights and measuring instruments can be checked and adjusted quickly and at favourable prices in KERN's accredited DAkkS laboratory (traceability to the national standard).
Adjustment instruction FK 1. Stable external working conditions shall be provided. A short warm-up phase of the device of approx. 1 minute is recommended for stabilisation. The FK can be switched on by pressing the "ON/OFF" key. The unit is in a hanging position.
Page 10
4. The corresponding weight is now attached to the hook. The appliance must be held still while the weight is attached; avoid shaking. Tip: hold the device with both hands while both elbows are resting on a table or attach it to a solid object (test bench).