14.2 Plug-in power supply unit
In some countries, a different plug-in power supply unit may be required for connection
to the local mains supply. The adapter or plug-in power supply used must not exceed
the output values at the USB output:
-
Voltage: 5V DC
-
Current: 1A
Do not use a defective or damaged plug-in power supply! There is a danger
to life due to high voltage!
14.3 Power supply via PC
The unit can also be powered from a PC using a USB C cable to simultaneously
transfer data via the USB interface.
15 Interface
15.1 Interface description
15.1.1 USB
This is a virtual USB interface. The device is not recognised as a storage medium!
Baud rate: 115200 Databits: 8
15.2 Interface protocol
The interface protocol KCP (KERN Communication Protocol) can be downloaded from
the website
www.kern-sohn.com.
Important ASCII interface commands:
-
SI
Sends the current reading immediately without waiting for stable
conditions.
-
SIR
Send current display immediately and repeat
-
U
Query of the display unit
-
ZI
immediately zero
-
@
cancel / unsubscribe
16 Declaration of conformity
To view the CE Declaration, please click on the following link:
https://www.kern-sohn.com/shop/de/DOWNLOADS/
28
Stopbits:1
Parity: none
FS-BA-en-2210.docx