Table of Contents

Advertisement

11.2.3.3 Manual without statistics
Use this mode to manually save the measured values currently displayed after
initialisation at the touch of a button.
1. Press the START/STOP button. The display shows "Manual running". This
value is not yet saved.
2. Each additional short press of the START/STOP button saves the currently
displayed readings. The number of stored values is displayed at the bottom
right. To the left is the file name in which the data is saved.
3. Press and hold START/STOP until the display shows "Manually ended" to end
the measurement series. No further points can be saved in this file.
4. By pressing the START/STOP button again, the mode starts from the beginning
and a new file is created.
11.2.3.4 Manual with statistics
The manual storage of the measured values is the same as described in chapter
11.2.3.3 described.
In addition, the min, max and average values of the stored measured values are
displayed here.

11.2.4 Channel limits

This setting controls the traffic light displayed when you are working in the
Measurement window. Change the values by clicking on them with the touch pen.
If the function is not required, enter the negative value of the sensor as MIN and the
positive value as MAX. The traffic light remains permanently yellow.
11.2.4.1 Yellow
Displayed measured value is below the entered value MIN
11.2.4.2 Green
Displayed value is between the entered values MIN and MAX
11.2.4.3 Red
Displayed measured value is above the entered value MAX
FS-BA-en-2210.docx
21

Advertisement

Table of Contents
loading

Table of Contents