Accuracy Table; Información Para Pedidos - BRAND Transferpette-8 electronic Operating Manual

Micropipettes
Hide thumbs Also See for Transferpette-8 electronic:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 48

7 Accuracy table

*) = Calculation for accuracy (A%) coefficient of variation (CV%): A% and CV% are calculated using the
formulas of statistical quality control.
Test instructions (SOPs) are available for download at www.brand.de .
7 Accuracy table
Volume
Partial
range
volume
[μl]
[µl]
0.5 - 10
10
  5
  1
1 - 20
20
10
  2
5 - 100
100
  50
  10
10 - 200
200
100
  20
15 - 300
300
150
  30
50 - 1250
1250
  625
  125
*A = Accuracy, CV = Coefficient of Variation
20 °C
!
Ex
Final test values based on the nominal volume (= max. volume) printed on the device and the speci-
fied partial volumes at the same temperature (20 °C/68 °F) of the device, surroundings and distilled
water, in accordance with DIN EN ISO 8655.
68
Operating manual
NOTICE
A* ≤ ± %
CV* ≤ %
1.2
0.8
2.0
1.5
8.0
4.0
1.0
0.5
2.0
1.0
8.0
3.0
0.8
0.25
1.6
0.4
4.0
1.5
0.8
0.25
1.4
0.4
4.0
1.3
0.6
0.25
1.2
0.4
3.0
1.2
0.8
0.25
1.2
0.4
4
1.2
Sub
Recommended tip type
steps
[μl]
[μl]
0.01
0.5 - 20
0.02
0.5 - 20
0.1
2 - 200
0.2
2 - 200
0.5
5 - 300
1.0
50 - 1250
Transferpette® -8/-12 electronic 997200

Advertisement

Table of Contents
loading

Table of Contents