Final test values related to the nominal capacity (maximum volume) indi-
cated on the instrument, obtained when instrument and distilled water
are equilibrated at ambient temperature (20 °C/68 °F) and with smooth
operation. According to DIN EN ISO 8655.
Accuracy tolerances for the Transferpette
Volume range
Volume step
µl
µl
30 - 300
10 - 100
2 -
20
A* = Accuracy, CV* = Coefficient of Variation
®
-8/-12 electronic
A*
≤ ≤ ≤ ≤ ≤ ± %
300
0.6
150
1.2
30
3.0
100
0.8
50
1.6
10
4.0
20
1.0
10
2.0
2
8.0

Accuracy Table

CV**
Increment
≤ ≤ ≤ ≤ ≤ %
µl
0.25
0.2
0.4
1.2
0.25
0.2
0.4
1.5
0.5
0.02
1.0
3.0
20 °C
!
Ex
Type of tips
µl
300
200 / 300
20
55