8 Adjustment – Easy Calibration

8.1 Adjustment

a.
1 x
> 3 s
b.
+
-
c.
1 x
d.
+
-
e.
1 x
Transferpette® -8/-12 electronic 997200
The nominal volume or volume to be tested is set, standard
mode Pipetting (PIP), e.g. 200.0 µl (for procedure, see PIP
mode, p. 58). Ex: Volume according to volume test 201.3 µl.
a.
Bring up CAL mode: Press and hold the MENU key (> 3
sec) until CAL appears.
ð
The display reads ‚off'.
ð
'CAL' blinks.
b.
Activate CAL mode: Press one of the arrow buttons to
activate the CAL mode.
ð
The display changes from ‚off' to ‚on'.
ð
'CAL' continues to blink.
c.
Confirm CAL mode: Press the Enter key.
ð
The display now shows the set pipetting volume.
ð
'CAL' blinks.
d.
Setting the volume: Use the arrow keys (+/-) to set the
volume, which was previously determined and tested.
ð
'CAL' blinks.
e.
Confirm volume: Press the Enter key.
ð
The display shows the tested and corrected volume.
ð
The CAL symbol is continuously displayed to confirm
that an adjustment has been made.
8 Adjustment – Easy Calibration
Operating manual
69