5.4 Electrophoresis (GEL) mode

Program for loading electrophoresis gels. A previously defined sample volume is aspirated at a high,
variable speed and slowly released again.
Volume and speed setting, see Set volume, p. 54 and Set the aspiration and dispensing speed, p. 55.
62
Operating manual
e.
Aspirate liquid: Press the pipetting key once to aspirate
the liquid.
When aspirating the liquid, slightly more volume is aspirated
than set.
f.
Dispense liquid in revPIP mode: Press the pipetting key
once to dispense. On the display, the arrow points
down (dispensing). The set volume is now dispensed
and some liquid remains in the tip.
g.
Re-aspirate liquid in revPIP mode: Pressing the pipet-
ting key again now resumes the set volume. (Pressing
the pipetting key again releases the volume)
h.
Trigger blow-out: After the last pipetting, press 'Enter'
key.
ð
The display now once again shows 'blo' for blow-out.
i.
End pipetting: By pressing the pipetting key once, the
blow-out is triggered, and the residual liquid is dis-
pensed.
ð
After dispensing the residual liquid (blow-out), the dis-
play jumps back to the set mode (start position).
NOTICE
Transferpette® -8/-12 electronic 997200