7 Accuracy table

Volume
Partial
range
volume
[μl]
[µl]
0.5–10
10
  5
  1
2 - 20
20
10
  2
10 - 200
200
100
  20
50 - 1000
1000
  500
  100
250 - 5000
5000
2500
  500
*A = Accuracy, CV = Coefficient of Variation
20 °C
!
Ex
Final test values based on the nominal volume (= max. volume) printed on the device and the speci-
fied partial volumes at the same temperature (20 °C/68 °F) of the device, surroundings and distilled
water, in accordance with DIN EN ISO 8655.
Transferpette® electronic 997209
NOTICE
A* ≤ ± %
CV* ≤ %
1.0
0.4
1.5
0.8
5.0
2.0
1.0
0.4
1.5
0.8
5.0
2.5
0.8
0.2
1.2
0.3
4.0
0.6
0.6
0.2
1.0
0.3
3.0
0.6
0.6
0.2
1.0
0.3
3.0
0.6
Sub
Recommended tip type
steps
[μl]
[μl]
0.01
0.5 - 20
0.02
0.5 - 20
0.2
2 - 200
1.0
50 - 1000
5.0
500 - 5000
7 Accuracy table
65