Technical Data; Accuracy Table - BRAND Handystep S Operating Manual

Repetitive pipettes
Hide thumbs Also See for Handystep S:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 32

9 Technical data

9 Technical data

9.1 Accuracy table

PD-Tips II from BRAND, 20 °C 'Ex'
PD-Tips
sizes
0.1 ml
0.5 ml
1.0 ml
1.25 ml
2.5 ml
5.0 ml
10.0 ml
12.5 ml
25.0 ml
50.0 ml
A* = Accuracy, CV = Coefficient of Variation
Error limits in relation to the set dispensing volume depending on the PD-
Tip size, at equal temperature (20 ºC) of instrument, tip, ambient environ-
ment and distilled water, and with uniform, jolt-free handling. The test is
carried out in accordance with DIN EN ISO 8655-5. The nominal volume is
the maximum volume printed on the PD-Tip.
Operating manual
52
Volume range
A* ≤ ± %
(µl)
stroke setting % of nominal
volume
1   2%
2 - 10
4.0
10 - 50
2.5
20 - 100
2.5
25 - 125
2.5
50 - 250
1.8
100 - 500
1.8
200 - 1000
1.8
250 - 1250
1.8
500 - 2500
1.5
1000 - 5000
1.5
CV* ≤ %
stroke setting % of nominal
volume
3   6%
5   10%
2.4
1.6
6.0
1.5
1.0
2.5
1.5
1.0
2.0
1.4
0.9
2.0
1.1
0.7
1.5
1.1
0.7
1.5
1.1
0.7
2.0
1.1
0.8
3.2
0.9
0.6
3.0
0.8
0.5
5.0
1   2%
3   6%
5   10%
3.0
2.0
1.5
1.0
1.2
0.8
1.1
0.7
0.9
0.6
0.9
0.7
1.2
0.8
2.0
1.4
1.5
1.0
1.8
1.2
997463