Accuracy Table - BRAND Transferpette S F-10 Operating Manual

Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 21

Accuracy Table

Transferpette
Volume range
µl
0.1 - 1
0.5 - 10
2 - 20
10 - 100
20 - 200
100 - 1000
500 - 5000
1000 - 10000
Transferpette
Volume range
µl
10
20
25
50
100
200
500
1000
28
®
S , Digital adjustable
Volume step
µl
1
0.5
0.1
10
5
1
20
10
2
100
50
10
200
100
20
1000
500
100
5000
2500
500
10000
5000
1000
®
S , Fixed volume
A*
CV*
≤ ± %
≤ %
1
0.5
0.8
0.4
0.8
0.4
0.8
0.4
0.6
0.2
0.6
0.2
0.6
0.2
0.6
0.2
A*
CV*
Increment
≤ ± %
≤ %
µl
2
1.2
0.001
4
2.4
20
12
1
0.5
0.01
1.6
1
7
4
0.8
0.4
0.02
1.2
0.7
5
2
0.6
0.2
0.1
0.8
0.4
3
1
0.6
0.2
0.2
0.8
0.3
3
0.6
0.6
0.2
1
0.8
0.3
3
0.6
0.6
0.2
5
0.8
0.3
3
0.6
0.6
0.2
10
0.8
0.3
3
0.6
Type of tip
µl
20
200
200
200
200
200
1000
1000
Type of tip
µl
20
nano-cap
20
200
200
200
1000
5000
10000
20 °C
!
Ex
Final test values related
to the nominal capacity
(maximum volume) indi-
cated on the instrument,
obtained when instrument
and distilled water are
equilibrated
at
ambient
temperature (20 °C/68 °F)
and with smooth opera-
tion. According to DIN EN
ISO 8655.
* A = Accuracy, CV =
Coefficient of Variation

Advertisement

Table of Contents
loading

Table of Contents