7 Accuracy table

7 Accuracy table
Volume
range
[µl]
0.5 - 10
5 - 50
10 - 100
20 - 200
30 - 300
*A = Accuracy, CV = Coefficient of Variation
20 °C
!
Ex
Final test values based on the nominal volume (= max. volume) printed
on the instrument and the specified partial volumes at the same temper-
ature (20 °C/68 °F) of the instrument, surroundings and distilled water,
in accordance with DIN EN ISO 8655.
Operating manual
48
Partial
A*
volume
[µl]
10
1.6
5
2
1
8
50
0.8
25
1.4
5
6
100
0.8
50
1.4
10
4
200
0.8
100
1.4
20
4
300
0.6
150
1.2
30
3
Sub
± CV*
%
steps
[µl]
1.0
0.01
2
6
0.4
0.05
0.8
3
0.3
0.1
0.6
2
0.3
0.2
0.6
1.5
0.3
0.5
0.6
1.5
Transferpette® S-8/-12 9974 89 V1
Recom-
mended
tip type
[µl]
0.5 - 20
2 - 200
2 - 200
2 - 200
5 - 300