Technical Data
PD-Tips II from BRAND, 20 °C 'Ex'
PD-Tip
Volume
size
range (µl)
0.1 ml
2 -
0.5 ml
10 -
1.0 ml
20 - 100
1.25 ml
25 - 125
2.5 ml
50 - 250
5.0 ml
100 - 500
10.0 ml
200 - 1000
12.5 ml
250 - 1250
25.0 ml
500 - 2500
50.0 ml
1000 - 5000
A* = Accuracy, CV* = Coefficient of variation
Error limits refer to the dispensing volume set relative to the PD-Tip, obtained when instrument,
PD-Tip and distilled water are equilibrated at ambient temperature (20 °C/68 °F) and with
smooth operation. The testing is according to DIN EN ISO 8655-5.
The nominal volume is the maximum volume printed on the PD-Tip.
Note:
The device is marked in accordance
with the German Measurement and
Calibration Act as well as the Meas-
urement and Calibration Regulation:
19
Character string: DE-M (DE for
Germany), framed by a rectangle, as
well as the last two digits of the year
in which the marking was affixed
(here: 2019).
A*
Stroke setting
of nominal volume
1
2%
3
=
=
10
4.0
2.4
50
2.5
1.5
2.5
1.5
2.5
1.4
1.8
1.1
1.8
1.1
1.8
1.1
1.8
1.1
1.5
0.9
1.5
0.8

Accuracy Table

± %
<
%
=
6%
5
10%
1
=
=
1.6
6.0
1.0
2.5
1.0
2.0
0.9
2.0
0.7
1.5
0.7
1.5
0.7
2.0
0.8
3.2
0.6
3.0
0.5
5.0
Important!
The HandyStep
operated with compatible tips from
other manufacturers. The user
must check their compatibility and
perform a function and volume test
before use!
CV*
%
<
Stroke setting
%
=
of nominal volume
2%
3
6%
5
10%
=
=
3.0
2.0
1.5
1.0
1.2
0.8
1.1
0.7
0.9
0.6
0.9
0.7
1.2
0.8
2.0
1.4
1.5
1.0
1.8
1.2
S can also be
®