6 Accuracy table

6 Accuracy table
Transferpette
Volume
range
[µl]
0.1–1
0.1–2.5
0.5–10
2–20
5–50
10–100
20–200
100–1000
500–5000
1000–10,000
*A = Accuracy, CV = Coefficient of Variation
Operating manual
48
®
S, adjustable
Partial
*R* ≤ ± %
volume
[μl]
1
2
0.5
4
0.1
20
2.5
1.4
1.25
2.5
0.25
12
10
1
  5
1.6
  1
7
20
0.8
10
1.2
  2
5
50
0.8
25
1.2
  5
4
100
0.6
  50
0.8
  10
3
200
0.6
100
0.8
  20
3
1000
0.6
  500
0.8
  100
3
5000
0.6
2500
0.8
  500
3
10000
0.6
  5000
0.8
  1000
3
CV* ≤ %
Sub
steps
[μl]
1.2
0.001
2.4
12
0.7
0.002
1.5
6
0.5
0.01
1
4
0.4
0.02
0.7
2
0.3
0.05
0.5
2
0.2
0.1
0.4
1
0.2
0.2
0.3
0.6
0.2
1
0.3
0.6
0.2
5
0.3
0.6
0.2
10
0.3
0.6
Recommended tip
type
[μl]
0.1–20
0.5–20
0.5–20
2–200
2–200
2–200
2–200
50–1000
500–5000
1000–10000
Transferpette® S 997207