To Clean the EI Ion Source

Materials needed
• Abrasive paper (5061-5896)
• Alumina abrasive powder (8660-0791)
• Aluminum foil, clean
• Cloths, clean (05980-60051)
• Cotton swabs (5080-5400)
• Glass beakers, 500 mL
• Gloves, clean, lint-free
• Solvents
• Ultrasonic bath
Preparation
1
2
5975/5977 Series MSD Operation Manual
Large (8650-0030)
Small (8650-0029)
Acetone, reagent grade
Methanol, reagent grade
Methylene chloride, reagent grade
Disassemble the ion source. See
Ion Source"
on page 99 or
page 102.
Collect the following parts to be cleaned for a standard or inert EI ion
source: (See
Figure 15
Repeller
Interface socket
Source body
Drawout plate
Drawout cylinder
Ion focus lens
Entrance lens
"To Disassemble the Standard or Inert EI
"To Disassemble the Extractor EI Ion Source"
on page 107.)