Agilent Technologies 5975 MSD Operation Manual page 50

For openlab cds
Hide thumbs Also See for 5975 MSD:
Table of Contents

Advertisement

3
Operating in Electron Ionization (EI) Mode
To enable/disable LVDS on MSD
1
2
3
50
Press Config, and then press MS/Aux Det to bring up the CONFIGURE MS
DETECTOR menu.
Press the down arrow to scroll to Lvds communication.
If you want to enable the LVDS, press On/Yes.
If you want to disable the LVDS, press Off/No.
Wait for the MSD to confirm the change on the GC control panel.
5975/5977 Series MSD Operation Manual