Agilent Technologies 5975 MSD Operation Manual page 35

For openlab cds
Hide thumbs Also See for 5975 MSD:
Table of Contents

Advertisement

7 Position the column so it extends 4 to 6 mm past the end of the ferrule
Figure 4
8 Slide the septum up to the bottom of the nut to fix the correct column
9 Insert the column in the inlet.
10 Slide the nut up the column to the inlet base and finger-tighten the nut.
11 Adjust the column position so the septum is even with the bottom of the
12 Tighten the column nut an additional 1/4 to 1/2 turn. The column should
13 Start carrier gas flow.
14 Verify flow by submerging the free end of the column in isopropanol. Look
See also
For more information about installing a capillary column, refer to Optimizing
Splitless Injections on Your GC for High Performance MS Analysis, Agilent
Technologies publication number 5988-9944EN.
5977B Series MSD Operation Manual
(Figure
4).
Insulation cup
Reducing nut
Capillary column
Ferrule (inside nut)
Inlet column nut
Septum
Installing a capillary column for a split/splitless inlet
insertion length.
column nut.
not slide with a gentle tug.
for bubbles.
Installing GC Columns
4 to 6 mm
2
35

Advertisement

Table of Contents
loading

This manual is also suitable for:

5977 msd

Table of Contents