Precautions - Agilent Technologies 5975 MSD Operation Manual

For openlab cds
Hide thumbs Also See for 5975 MSD:
Table of Contents

Advertisement

1
Introduction
Table 4
Mechanism
Mass spectrometer manual shutoff
valves closed
GC off
Power failure
Once hydrogen has accumulated in a mass spectrometer, extreme caution must be
WARNING
used when removing it. Incorrect startup of a mass spectrometer filled with
hydrogen can cause an explosion.
After a power failure, the mass spectrometer may start up and begin the pumpdown
WARNING
process by itself. This does not guarantee that all hydrogen has been removed from
the system or that the explosion hazard has been removed.

Precautions

Take the following precautions when operating a GC/MSD system with
hydrogen carrier gas.
20
Hydrogen accumulation mechanisms (continued)
Results
Some mass spectrometers are equipped with manual
diffusion pump shutoff valves. In these instruments, the
operator can close the shutoff valves. Closing the shutoff
valves does not shut off the flow of carrier gas. As a result,
hydrogen may slowly accumulate in the mass spectrometer.
A GC can be shut down deliberately. It can also be shut
down accidentally by an internal or external failure.
Different GCs react in different ways. If a 7890 Series GC
equipped with Electronic Pressure Control (EPC) is shut off,
the EPC stops the flow of carrier gas. If a GC's carrier flow is
not under EPC control, the flow increases to its maximum.
This flow may be more than some mass spectrometers can
pump away, resulting in the accumulation of hydrogen in
the mass spectrometer. If the mass spectrometer is shut off
at the same time, the accumulation can be fairly rapid.
If the power fails, both the GC and mass spectrometer shut
down. The carrier gas, however, is not necessarily shut
down. As described previously, in some GCs, a power failure
may cause the carrier gas flow to be set to maximum. As a
result, hydrogen may accumulate in the mass spectrometer.
5975/5977 Series MSD Operation Manual