Gcms Transfer Line Tip Seals - Agilent Technologies 5975 MSD Operation Manual

For openlab cds
Hide thumbs Also See for 5975 MSD:
Table of Contents

Advertisement

13 Gently check the alignment of the ion source and the interface tip seal.
Forcing the analyzer door closed if these parts are misaligned will damage the seal or
CA UT IO N
the interface or the ion source, or will keep the sideplate from sealing.
14 You can align the ion source and interface tip seal by wiggling the side plate
15 Close the analyzer chamber door. (See

GCMS transfer line tip seals

Table 6
Model
Tip seal
Spring
Tip cap
Tip base
Table 7
Model
Tip seal
Tip cap
Tip base
5977B Series MSD Operation Manual
When the ion source is aligned correctly, the front analyzer chamber can be
closed all the way with no resistance except the spring tension from the
interface tip seal.
on its hinge. If the door still will not close, contact your Agilent
Technologies service representative.
page 125.)
EI GCMS interface transfer line tip seal for extractor ion sources
5977A
G3870-20542
G7005-20024
G3870-20543
------------------
CI GCMS transfer line tip seals for CI and extraction ion sources
5975
G1099-60412
-----------------
-----------------
"To Close the Analyzer Chamber"
5977B
G3870-20542
G1999-20023
G3870-20547
G3870-20548
5977A
5977B
G1999-60412
G1999-60412
G3870-20543
G3870-20547
------------------
G3870-20548
Installing GC Columns
2
on
41