To Disassemble The Extractor Ei Ion Source - Agilent Technologies 5975 MSD Operation Manual

For openlab cds
Hide thumbs Also See for 5975 MSD:
Table of Contents

Advertisement

4
General Maintenance

To Disassemble the Extractor EI Ion Source

Materials needed
• Gloves, clean, lint-free
• Hex ball driver, 1.5 mm (8710-1570)
• Hex ball driver, 2.0 mm (8710-1804)
• Wrench, open-end, 10 mm (8710-2353)
Procedure
1
2
3
4
5
6
7
8
102
Large (8650-0030)
Small (8650-0029)
Remove the ion source. See
Remove the filaments by removing the two gold plated screws and
separating the filaments from the source. See
Loosen the two gold plated screws on the source heater block assembly,
and separate the repeller assembly from the source body. The repeller
assembly includes the source heater block assembly, repeller, and related
parts.
Remove the gold plated setscrews from the side of the source body.
Pull the entrance lens and ion focus lens to remove them from the source
body.
Remove the extractor lens and insulator.
Separate the entrance lens and ion focus lens from the lens insulator.
Remove the repeller nut, washers, and insulators from the source heater
block assembly, then remove the repeller.
"To Remove the EI Ion Source"
Figure 14
5975/5977 Series MSD Operation Manual
on page 96.
on page 103.