To Clean the EI Ion Source

C A U T I O N
Troubleshooting and Maintenance Guide
Materials needed
• Abrasive paper (5061-5896)
• Alumina abrasive powder (8660-0791)
• Aluminum foil, clean
• Cloths, clean (05980-60051)
• Cotton swabs (5080-5400)
• Glass beakers, 500 mL
• Gloves, clean, lint-free
Large (8650-0030)
Small (8650-0029)
• Solvents
Acetone, reagent grade
Methanol, reagent grade
Methylene chloride, reagent grade
• Ultrasonic bath
Preparation
1
Disassemble the ion source. See
2
Collect the following parts to be cleaned:
Repeller
Interface socket
Source body
Drawout plate
Drawout cylinder
Ion focus lens
Entrance lens
These are the parts that contact the sample or ion beam. The
other parts normally should not require cleaning.
If insulators are dirty, clean them with a cotton swab dampened with
reagent-grade methanol. If that does not clean the insulators,
replace them. Do not abrasively or ultrasonically clean the
insulators.
General Maintenance
page
141.
(Figure
13)