To Clean the EI Ion Source
C A U T I O N
Troubleshooting and Maintenance Guide
Materials needed
• Abrasive paper (5061-5896)
• Alumina abrasive powder (8660-0791)
• Aluminum foil, clean
• Cloths, clean (05980-60051)
• Cotton swabs (5080-5400)
• Glass beakers, 500 mL
• Gloves, clean, lint-free
•
Large (8650-0030)
•
Small (8650-0029)
• Solvents
•
Acetone, reagent grade
•
Methanol, reagent grade
•
Methylene chloride, reagent grade
• Ultrasonic bath
Preparation
1
Disassemble the ion source. See
2
Collect the following parts to be cleaned:
•
Repeller
•
Interface socket
•
Source body
•
Drawout plate
•
Drawout cylinder
•
Ion focus lens
•
Entrance lens
These are the parts that contact the sample or ion beam. The
other parts normally should not require cleaning.
If insulators are dirty, clean them with a cotton swab dampened with
reagent-grade methanol. If that does not clean the insulators,
replace them. Do not abrasively or ultrasonically clean the
insulators.
General Maintenance
page
141.
(Figure
13)
10
143