Operating in Electron Ionization (EI) Mode
4
To view MSD vacuum pressure and Turbo speed/Foreline Pressure from the
GC control panel
1
2
To vent the MSD from the GC control panel
1
2
3
4
To pump down the MSD from the GC control panel
1
2
3
4
To view the firmware version of the MSD from the GC control panel
1
2
To view the serial number of the MSD from the GC control panel
1
2
48
Press Enter to apply the changes.
Press MS/Aux Det to display the 5977B MSD menu.
Press the down arrow to scroll to HiVac Pressure, or Turbo Speed % of full
/
Foreline Pressure.
With the MSD pumped down, press MS/Aux Det to bring up the 5977B MSD
menu.
Press the down arrow to scroll to Start MSD Vent?. (Press Off/No to cancel
the vent cycle and pump down the MS).
Press ON/Yes to start the venting cycle.
When prompted open the vent valve.
With the MSD vented, press MS/Aux Det to bring up the 5977B MSD menu.
Press the down arrow to scroll to Start MSD Vent?.
Press ON/Yes to start the pump down cycle.
When prompted open the vent valve.
Press MS/Aux Det to bring up the 5977B MSD menu.
Press the down arrow to scroll to Firmware.
Press MS/Aux Det to bring up the 5977B MSD menu.
Press the down arrow to scroll to Serial#.
5975/5977 Series MSD Operation Manual