Dräger PSS BG 4 Manual page 69

Hide thumbs Also See for PSS BG 4:
Table of Contents

Advertisement

Testing
Symbol "Battery warning 2"
In this case the red LEDs flashes
constantly. On power up 5 short bleeps
will sound and the display will show the
"Battery warning 2" icon.
Bodyguard automatically powers off and
any further operation is impossible.
WARNING
If battery warning 2 is displayed, the
battery must be replaced before any
new operation.
Failure to follow this warning could
lead to death or serious injury.
Releasing Air from the
Breathing Circuit
Remove sealing cap R 33 588 (1).
Wait until Bodyguard shows 0 bar
pressure.
Place the sealing cap R 33 588 (1)
on the plug-in coupling.
Switching off Bodyguard
Simultaneously press the right and
left hand button until a single sharp
audible "Bleep" sounds.
Release buttons.
For 3 seconds Bodyguard shows the
battery status.
Bodyguard switches off.
5x Bleep
bar
1
bar
1x Bleep
bar
Проверка
Символ "Предупреждение о
разряде батареи 2"
В этом случае красные светодиоды
непрерывно мигают. При включении
подается 5-кратный короткий
звуковой сигнал и на дисплей
выводится значок "Предупреждение о
разряде батареи 2".
Bodyguard автоматически
выключается, и любые дальнейшие
операции невозможны.
ОСТОРОЖНО
Если показано второе
предупреждение о разряде батареи,
ее необходимо заменить до
продолжения эксплуатации
аппарата.
Игнорирование данного
предупреждения может привести к
гибели или серьезной травме.
Выпуск воздуха из
дыхательного контура
Снимите заглушку R 33 588 (1).
Ждите, пока Bodyguard не покажет
давление 0 бар.
Снова установите заглушку
R 33 588 (1) на штекерный
соединитель.
Выключение Bodyguard
Одновременно нажмите и
удерживайте правую и левую
кнопки, пока прибор не подаст
однократный резкий звуковой
сигнал.
Отпустите кнопки.
В течение 3-х секунд Bodyguard
показывает состояние батареи.
Bodyguard выключается.
69

Advertisement

Table of Contents
loading

Table of Contents