Electrodes should not be placed over muscles, because muscle trembling causes too much
noise on the record.
Patient must remain still during record creation, he should not move or talk.
In case you are using disposable electrodes, you should check the expiry date of the
electrodes.
7.1.1. Limb electrode placement
R
N
7.1.2. Chest electrode placement
18/61
HS80G-L1 Users' Manual
R:
L
L:
N:
F:
F
V1 : on the IVth rib space on the right side of the
sternum
V2: on the IVth rib space on the left side of the
sternum
V3: on the midpoint between the line, connecting
the leads V2 to V4
V4: on the Vth rib space on the mediclavicular
line
V5: on the Vth rib space on the frontal armpit
line
V6: on the Vth rib space on the central armpit
line
right hand
left hand
right leg
left leg
R-7317-2-B