Now you can view the record and the related data, according to the descriptions of the
Section, titled Handling of stored data, or you can return to monitoring.

9.1.3. Aborting the program

Program operation can be aborted by pressing the
key, but in such case no storage is made.
Making analysis-diagnosis – the DIAG program
9.2.
The analysis / diagnosis program collects 16-second long section from the 12 ECG channels,
performs R-wave averaging, calculates amplitude and time parameters, and then creates a
diagnosis recommendation based on the following data:
 a typical averaged QRS cycle from all the 12 channels
 analysis data from all the 12 channels
 a 16-second long 1-channel rhythm record.
The rhythm record, which the DIAG program stores, only shows the possible
arrhythmias and only 5mm/s printing speed is available.
The 12 typical – so called averaged QRS – complexes are only evaluated based on the typical
beats, the non-characteristic QRS's are filtered. Their lengths are always 880ms.
Warning!
The device can only create analysis in the 45-175 heart rate range.
If pre-start ('Settings / Printing / Pre-start') is enabled, the 2-second long signal-section prior
to the push of the button is also stored. This makes it possible to store the curve already
displayed.
R-7317-2-B
HS80G-L1 Users' Manual