 year, month, day
 month, day, year
 day, month, year.
(Default: month, day, year)
The adjustments can be saved with the ESC or the
.
menu level

13.7. Filters

1:
Here you can set the mains frequency for the mains filter. The mains frequency can be
set to either 50 or 60 Hz (Europe, America). (50 Hz)
2:
Operation of the baseline filter can be enabled. The baseline filter offers the creation of
better quality records by suppressing baseline movement. During monitoring and
printing the baseline filter ensures that the low frequency noise on the electrodes, noise
from patient breathing and most of the noise from patient movement do not affect the
ECG signal. (on)
13.8. HR Warning
The purpose of HR Warning is for the operator to receive a sign, if the HR exceeds a limit.
The upper and lower limit of the range can be set. The signal consists of audio and visual
signs with a continuous alarming sound and with a ringing bell on the display.
1:
The warning function can be set on/off. (off)
2:
Heart rate upper limit. Adjustable between 60 and 140 in steps of 5 (120).
3:
Heart rate lower limit. Adjustable between 30 and 80 in steps of 5 (40).
13.9. Other
1:
QRS sound can be turned on/off. (on)
2:
Sound of the keyboard can be set on/off in this menu item. (off)
3:
The device has a Bluetooth module which helps sending records to the PC for archiving
and ensures online monitoring on the PC. Bluetooth can be turned on/off in this line
(off). If the module is turned on, the identifier of the Bluetooth module can be seen on
the bottom line of the Menu screen. (e.g.: HS80CL-FD12, in which case FD12 is the
identifier.)
You can enable or disable the device's automatic shutdown service. If enabled, the
4:
device automatically shuts down after 5 minutes, when none of the controls are used
and there is no continuous PC connection (5 mins).
5:
Lead groups can be standard or Cabrera. (standard)
6:
Memory erase. This is a complete reset, return to default values. All stored records are
deleted, the automatic counter of patient identifier is reset to 0001 and the settings
return to their default values.
50/61
HS80G-L1 Users' Manual
,
button
with that returning to the previous
R-7317-2-B