Role Of The Touch-Screen During Monitoring, Freeze; Operation Of The Touch-Screen In Menu Mode; Useful Pieces Of Advice On How To Make A Good Quality Ecg Record; How To Make A Good Ecg Record - Innomed Medical HeartScreen 80G-L1 User Manual

Table of Contents

Advertisement

Touching the last two crosses is already for the checking of the calibration. If the calibration
is faulty, namely the device detects as if it was not the cross that was touched on the display,
it will not step forward. In such case, calibration can be re-started by touching the zone of
different color.

6.1.2. Role of the touch-screen during monitoring, freeze

If device is turned on, after the self-test it enters monitoring mode, three ECG channels will
be displayed on the screen at a speed of 25 mm/s and with 10 mm/mV sensitivity. At the
bottom of the screen, a rhythm curve appears, at 5 mm/s speed. If the surface of the display is
touched, on that part where the 3 ECG curves are running, the image will be freezed. (freeze
function). For another touch, displaying of the monitored curves will continue. The same way
rhythm curve can be freezed as well, if the screen is touched around the zone of the curve.
Freezing can be terminated by touching the screen again.

6.1.3. Operation of the touch-screen in menu mode

Operation of the touch-screen in menu modes will be detailed under the description of menu
modes, as operation can only be understood together with the operation mode description.
General remark about the handling: in case, in the description, we are talking about a key
displayed on the screen, its using (touching) is to be understood as follows: the named key is
touched on the screen, or the key on the keyboard, next to the displayed key is pressed.
Useful pieces of advice on how to make a good quality
7.
ECG record
7.1.

How to make a good ECG record

INNOMED ships its devices with silver-silver chloride electrodes.
These electrodes have low contact potentials, allowing the creation of much better quality
ECG than with non-silver-silver chloride electrodes.
Please use our devices with these electrodes.
Connect the patient cable (4 or 10-lead) to the (16) connector, located at the back side of
the device (figure no. 15). Connect the connectors on the other side of the cable to the
electrodes, according to the figure below.
Oil and body lotion should be removed from patient's skin by alcohol, if necessary,
patient should be shaved (for example, in case of hairy chest).
Electrodes should also be cleaned before each use. Always make sure that no dried gel
should remain on the electrodes!
INNOMED supplied gel must be applied on the electrodes (or similar quality gel
recommended for ECG).
The use of wet gauze instead of gel is NOT recommended.
R-7317-2-B