HS80G-L1 Users' Manual

Computer connection

14.
The Innobase for Windows PC-based cardiological database management program can be
purchased for the HeartScreen 80G-L1. The program is responsible for patient database
management, measurement result storage, monitoring on 12 channels, measurement result
display, evaluation, and size A4 printing. Use of the program can be learned from its own
manual.
The device can be connected to the computer via USB cable or with Bluetooth connection.
When using USB cable, turn off the Bluetooth module in the Settings/Other menu item – if it
was turned on previously. Connect the device to the PC with the appropriate cable (plug (15)
in the figure on page 15), turn the device on and start the Innobase for Windows program on
the PC. The program connects to the device and from this point all control is from the PC,
where all functions (sensitivity, speed, filter setting, monitoring, printing, starting
programmed measurements, etc.) are available.
When using Bluetooth, turn on the Bluetooth module in the Settings/Other menu item. An
installed Bluetooth dongle shall be on the computer matched with HeartScreen 80G-L1
according to its own manual. After starting the Innobase for Windows program, the PC
connects to the device and from this point all control is from the PC, where all functions
(sensitivity, speed, filter setting, monitoring, printing, starting programmed measurements,
etc.) are available.