Heart Screen 80G-Lprinting And Display Properties; Heartscreen 60G Printing And Display Properties; Sio Parameters; Mechanical Data (Informative) - Innomed Medical Heart Screen 60G Service Manual

Simultaneous 12-channel ecg devices with 58 / 80mm wide 1- / 3-channel built-in printer and lcd display
Hide thumbs Also See for Heart Screen 60G:
Table of Contents

Advertisement

Innomed Medical
Data acquisition
Sampling frequency
A/D resolution
Referring to ± 10 mV signal range
Stored signal resolution
Serial line resolution during on-line ECG transfer
4.10
HeartScreen 80G-L printing and display properties
Effective recording width per channel
Printer head resolution
Paper speed
Max. paper run speed
(in case of 60bpm, informative)
Paper width
Paper roll outer diameter
Printer lifetime
Paper change
Display
4.11

HeartScreen 60G printing and display properties

Printer head resolution
Paper speed
Paper width
Paper roll outer diameter
Printer lifetime
Paper change
Display
4.12

SIO parameters

Serial I/O
Serial line services
4.13

Mechanical data (informative)

Size
Device weight with battery pack and charger
4.14

Pacemaker pulse detection

When the device examines a patient with a pacemaker, it recognizes the pacemaker signal within a set range
according to technical data and displays them graphically on the record. In this case, the device does not create a
diagnosis, since the pacemaker pulse interferes with the diagnosis program.
In this case the device only creates analysis, which the pace impulse does not interfere with, but due
!
to morphological differences there may be inaccuracies in the numerical data.
4.15

Checking electrode and amplifier operation

If the input side of the amplifier does not work properly or there is cable or electrode problem, the device blocks
automatically and initializes the filters. After blocking the device checks if the error state continues. If so, the
device repeats blocking and filter initialization.
2014.01.09
Heart Screen 60G-80G-L
12 leads, simultaneous
1000 Hz / channel
13 bit
2.5 µV/bit
500 Hz
10 µV/bit
500 Hz
10 µV/bit
40 mm, overlapping allowed
in amplitude direction: 8 points / mm
on time axis: 8 or 16 points / mm
5, 25,50 mm/s
25 mm/s
80 mm +0; -1%
max. 50 mm
50,000 m
automatic change
240 x 320 points
in amplitude direction: 8 points / mm
on time axis: 8 or 16 points / mm
5, 25,50 mm/s
58 mm +0; -1%
max. 50 mm
50,000 m
automatic change
64 x 128 points
RS – 232
on-line monitoring from the 12 channels
ECG curve archiving on the PC
8/84
250 * 150 * 70 mm
1.3kg