Setting Of Device And Patient Data, Access To Stored Records; Programs Of The Device; Record Storage - The Store Program; Starting The Program - Innomed Medical HeartScreen 80G-L1 User Manual

Table of Contents

Advertisement

Programs are detailed in the following section.

8.6.9. Setting of device and patient data, access to stored records

We can enter the stored records, device settings and patient identification menu by pressing
the
key. Guidelines about the handling of stored records, device settings and the patient
identification code and name settings can be found in the posterior sections.

Programs of the device

9.
Record storage – the STORE program
9.1.
This program makes adjustable length records from the 12 ECG leads, the Spec group;
adjustable length rhythm record from one of the channels, and if possible performs rhythm
analysis as well:
 0, 4, 6 or 8-second long record can be made of the ECG leads.
 0, 4, 6 or 8-second long simultaneous record can be made of the Spec. group.
 The length of the rhythm record made during the recording can be 0, 20 seconds, 1
minute, 3 minutes, 5 minutes or 10 minutes. Rhythm record can be made of any
channel.
The rhythm record, which the program stores, only shows the possible arrhythmias and
only prints at 5mm/s speed.
 If the preset rhythm record is 1 minute or longer, the device, besides the rhythm record,
also performs the rhythm analysis, if during the period of the rhythm record at least 40
beats occur. The printed format of the rhythm analysis is identical to the one specified
regarding the rhythm analysis program, to be mentioned later.
Storing parameters can be modified in the 'Settings/Storage' menu point.
If pre-start ('Settings / Printing / Pre-start') is enabled, the 2-second long signal-section prior
to the push of the button is also stored. This makes the storage of the already displayed curve
possible.

9.1.1. Starting the program

If you wish to make rhythm analysis and storage, you have to take the following steps first:
 Set the STORE program ECG and SPEC time to 0, the rhythm time to at least 1 minute
(Settings / Storage).
 There is no need for chest electrodes for this examination, so it is recommended to short
these by connecting to the N lead. When using chest electrodes the unstable chest
electrode contact may cause problems during several-minute (5-10 minutes)
examinations, automatic blocking may happen and a delay occurs in the record
regardless of appropriate signal from the limb leads. It is recommended to use the type
R-7317-2-B
HS80G-L1 Users' Manual
27/61