Monitoring mode

8.
After being switched on by the (1) ON/OFF key, in case of existing stored records, the device
performs a memory test, which is indicated by the
message, the displayed Innomed logo and the version number. In case of several stored
records, it might even take10-15 seconds.
After this, the device automatically, or in cases mentioned later by pressing the
enters monitoring mode.
This is the device starting, standard operating mode. Three selected ECG curves appear on the
display, together with the current pulse rate, lead, speed, sensitivity and filter state.
The device calls attention on possibly discharged batteries and the presence of pacemaker
impulse.
After power on, monitoring always starts with the same settings. These are the following:
leads: I, II, III; sensitivity: x1; speed: 25 mm/s; filters: off.
The display during monitoring
Érzékenység
Sensitivity
Élő Bluetooth
Bluetooth
connection
kapcsolat
Pacemaker
Pacemaker
Jelzés
indicator
Akkumulátor
Low battery
lemerült jelzés
indicator
Elvezetések
Lead
jelzése
indicator
Ritmus csatorna
Rhythm channel
elvezetés
lead
 Live Bluetooth connection indicates, that the device is in active connection with a PC
with the purpose of monitoring or data transfer.
 The "P" pacemaker indication appears, if the examined patient has implanted
pacemaker, and the device detects this. In this case, location of the pacemaker impulse
on the ECG curve is indicated by a marker.
 The "Battery low" indication appears if the device battery is lower than the 10 % of its
capacity. In such case, you should start battery charging.
 Beat type is defined according to the rhythmicity and the QRS width. There are
rhythmical, early and late beats and according to the width: wide, normal and extra
wide beats. Exact definition is described in the Section of Rhythm analysis.
Meaning of the rest of the information displayed on the screen is obvious.