Innomed Medical HeartScreen 80G-L1 User Manual page 24

Table of Contents

Advertisement

 V1, V2, V3
 V4, V5, V6
 Selected leads within the Spec. group
If pre-start (Settings / Printing / Pre-start) is enabled, the 2-second signal section prior to the
button push is printed. This makes it possible to print the already displayed.
During printing you can change the followings:
 lead groups,
 speed
 sensitivity
In such case printing re-starts with the validating sign.
In case of filter change, printer starts with a new header.
A typical printout:
As you can also see in the above image, the header of the record always includes the
followings:
 identification number, (if it is not given, the Auto xxxx automatically generated
identification number will be written here, where xxxx is a number of four characters,
created by the device.)
 if patient name is not given (there will be a space in its place)
 time and date of printing,
 paper speed and
 sensitivity,
 filter settings,
 type of the device and software version number.
Warning!
The automatic patient identifier only increases in case of storage or after starting the analysis /
diagnosis program, not when printing.
Following the header, the 300ms long, 1mV amplitude calibrating square signal is printed.
The 3 channels' curve only appears after these.
24/61
HS80G-L1 Users' Manual
R-7317-2-B