V1, V2, V3
V4, V5, V6
Selected leads within the Spec. group
If pre-start (Settings / Printing / Pre-start) is enabled, the 2-second signal section prior to the
button push is printed. This makes it possible to print the already displayed.
During printing you can change the followings:
lead groups,
speed
sensitivity
In such case printing re-starts with the validating sign.
In case of filter change, printer starts with a new header.
A typical printout:
As you can also see in the above image, the header of the record always includes the
followings:
identification number, (if it is not given, the Auto xxxx automatically generated
identification number will be written here, where xxxx is a number of four characters,
created by the device.)
if patient name is not given (there will be a space in its place)
time and date of printing,
paper speed and
sensitivity,
filter settings,
type of the device and software version number.
Warning!
The automatic patient identifier only increases in case of storage or after starting the analysis /
diagnosis program, not when printing.
Following the header, the 300ms long, 1mV amplitude calibrating square signal is printed.
The 3 channels' curve only appears after these.