 When using the KPAC4 type, 4-lead cable, the V1–V6 channels operate with shorted
inputs, and measurement preparation is also simple, only the limb clips need to be
placed.
 For monitored lead, select a limb lead, e.g. the II. lead.
Warning!
Warning! When using shorted, 10-lead cable with chest electrodes connected to the N
lead, noisy signal similar to ECG signal is visible on channels V1-V6, which is to be
ignored, so this signal cannot be associated with any standard lead! The device
automatically recognizes the type KPAC4 cable and leads V1...V6 do not appear
when used.
 Check whether there is enough paper in the device.
During monitoring, press the

9.3.2. Running the program

When starting the program, the program name appears on the top in the middle of the display
and next to it a counter, indicating in seconds the time left from the analysis. The maximum
duration of the analysis is 5 minutes, that is 300 seconds. Data acquisition process can be
interrupted, but for successful analysis and statistics creation the QRS detector should identify
minimum 40 beats.
After printing the header, in case of the above listed arrhythmias, the printer prints the 2-2
seconds before and after the event. If more abnormal beats happen during the printing of this
4-second section, they are not printed but appear in the statistics.
The program indicates the beat classification by the letters N, B, W, S, E or L above the
QRSs:
Warning!
Arrhythmia analysis and printing is based on the selected, monitored lead group.
Picture of printout, in case of printing abnormal events:
34/61
HS80G-L1 Users' Manual
key to start the selected program with header printing.
R-7317-2-B