Waters 3100 Operator's Manual page 47

Table of Contents

Advertisement

Caution:
enclosure door while the door is open.
3.
Open the source enclosure door by releasing both spring-clips and
lowering the door towards you.
4.
Use the needle-nose pliers to remove the blanking plug from the corona
pin mounting contact. Store the blanking plug in a safe location.
Corona pin mounting contact
Warning:
handle the corona pin with care.
Caution:
the pin, use the needle-nose pliers to grip the corona pin at the
end that fits into the mounting contact.
5.
Use the needle-nose pliers to fit the corona pin to the mounting contact.
Ensure that the corona pin is securely mounted and that
Requirement:
its tip aligns with the sample cone aperture.
Do not apply any downward force to the source
The corona pin tip is sharp. To avoid puncture wounds,
To avoid damaging to the corona pin's tip and bending
RP00028
RP00028
Corona pin mounting
contact blanking plug
Installing the corona pin
3-7