25. Attach the coupling nut to the probe, and gently tighten it with the
7/16-inch wrench.
Coupling
LC union
26. Replace the probe tip, and then screw down until the capillary protrudes
approximately 0.5 mm from the end of the tip.
27. Fit the probe end-cover.
28. Fit and tighten the two probe end-cover retaining screws.
29. Tighten the 3 set screws to clamp the LC union in place.
30. Fit the probe to the instrument (see
page
3-2).
Capillary protrudes 0.5 mm
from the end of the probe tip
"Installing the ESI probe" on
Replacing the ESI probe sample capillary
5-59