Waters 3100 Operator's Manual page 131

Table of Contents

Advertisement

Warning:
biohazardous and/or toxic materials. Dispose of it according
to local environmental regulations.
3.
Dispose of the probe heater in accordance with local environmental
regulations.
4.
Carefully slide the replacement probe heater over the capillary sleeve on
the probe assembly
Probe heater connections
Caution:
connections, do not twist the heater when fitting it to the probe
assembly.
5.
Fit the probe heater to the probe assembly, ensuring that the heater is
fully seated on the probe assembly.
6.
Fit the probe to the instrument (see
probe" on page
The probe heater can be contaminated with
To avoid damaging the probe heater's electrical
4-3).
Capillary sleeve
Capillary
"Installing the IonSABRE APCI
Replacing the APCI probe heater
5-71