Waters 3100 Operator's Manual page 108

Table of Contents

Advertisement

Caution:
ensure that the bottom of the assembly is not in contact with the
bottom of the glass vessel.
3.
Use the hook to carefully suspend the hexapole assembly into the glass
vessel with the bottom of the assembly clear of the bottom of the vessel.
4.
Add 1:1 methanol/water to the glass vessel until the hexapole assembly
is immersed completely.
5.
Place the vessel in the ultrasonic bath for 30 minutes.
Caution:
clean, chemical-resistant, powder-free gloves for the rest of this
procedure.
6.
Carefully remove the hexapole assembly from the vessel, and blow-dry it
using inert, oil-free gas.
5-48
Maintenance Procedures
To avoid vibration damage to the hexapole assembly,
Hook
Rear support ring
TP02658
To avoid recontaminating the hexapole assembly, wear