Waters 3100 Operator's Manual page 133

Table of Contents

Advertisement

3.
Use the 2.5-mm Allen wrench to loosen the 4 captive screws securing the
ion block cover plate.
4.
Remove the ion block cover plate.
5.
Use the 1.5-mm Allen wrench to remove the 2 screws securing the
heater wires to the PEEK terminal block.
Ion block cover plate captive
securing screw
Ion block cover plate
TP02649
TP02616
PEEK terminal block
Replacing the ion block source heater
Heater cartridge wire
securing screws
5-73