Waters 3100 Operator's Manual page 137

Table of Contents

Advertisement

To remove the source enclosure
Warning:
biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while performing this
procedure.
1.
Vent and shut-down the instrument (see the mass spectrometer's online
Help for
details).
Warning:
to cool for at least 30 minutes before proceeding.
2.
Remove the probe from the source.
If you are removing an ESI probe, see
page
3-10.
If you are removing an IonSABRE APCI probe, see
IonSABRE APCI probe" on page
Caution:
enclosure door while the door is open.
3.
Open the source enclosure door by releasing both spring-clips and
lowering the door towards you.
Warning:
handle the corona pin with care.
4.
If using ESCi mode or an IonSABRE APCI probe, carefully remove the
corona pin (see
5.
Disconnect the probe's electrical connection at the instrument's front
panel.
6.
Disconnect the PTFE tubing at the desolvation gas connection on the
front panel.
7.
Disconnect the 2-pin connector at the rear of the source enclosure.
The source components can be contaminated with
The source can be hot. To avoid burn injuries, allow it
Do not apply any downward force to the source
The corona pin tip is sharp. To avoid puncture wounds,
"Removing the corona pin" on page
Replacing the source assembly seals
"Removing the ESI probe" on
"Removing the
4-7.
3-8).
5-77