4.
Use the O-ring removal kit to carefully remove the seal from the source
enclosure door.
Door glass retaining
screw
Source enclosure
glass seal
Source enclosure
door seal
RP00019
5.
Use the screwdriver to remove the 4 screws securing the glass-retaining
plate to the source enclosure door.
6.
Remove the plate from the source enclosure door.
7.
Remove the glass from the source enclosure door.
8.
Use the O-ring removal kit to carefully remove the door glass seal from
the source enclosure door.
9.
Dispose of all the seals in accordance with local environmental
regulations.
5-82
Maintenance Procedures
The seals can be contaminated with biohazardous
Warning:
and/or toxic materials. Ensure that they are correctly
disposed of according to local environmental regulations.