To connect the roughing pump
Caution:
• To ensure correct operation of the roughing pump, the pump
• The area where the roughing pump is located must have an
• To ensure proper ventilation, the pump must be installed with
Left side minimum clearance
is 15.24 cm (6 inches)
Front side minimum clearance
is 35.56 cm (14 inches)
Requirement:
daily access to the gas ballast valve and oil-level sight glass.
1.
Place the PTFE drip tray on the floor, within 5 feet of the instrument.
Warning:
two people must lift the pump.
2.
Place the pump on the PTFE drip tray.
B-4
External Connections
The pump and its connections can be
Warning:
contaminated with biohazardous and/or toxic materials.
Always wear chemical-resistant, powder-free gloves when
performing this procedure.
must be installed within 1 degree of horizontal.
ambient temperature of 15 to 40 ºC (59 to 104 ºF).
the following minimum clearances:
The pump must be oriented in a way that allows easy
The roughing pump is heavy. To avoid injury, at least
Back side minimum clearance
is 15.24 cm (6 inches)
TP02689
Right side minimum clearance
is 15.24 cm (6 inches)