Waters 3100 Operator's Manual page 183

Table of Contents

Advertisement

Sharp knife
1-L exhaust trap bottle (included in the startup kit)
Elbows (included in the Waters Rough Pump Connect Kit)
NW25 tee (included in the startup kit)
NW25 center rings (included in the startup kit)
NW25 clamps (included in the startup kit)
NW25 full nipple (included in the Alcatel pump kit)
DN25 reducing nipple (included in the Alcatel pump kit)
NW40 center rings (included in the Alcatel pump kit)
NW40 clamps (included in the Alcatel pump kit)
DN40 full nipple flange (included in the Alcatel pump kit)
PVC exhaust tubing (included in the Waters Rough Pump Connect Kit)
PVC hose clamps (included in the Waters Rough Pump Connect Kit)
1-inch ID vacuum hose (included in the Waters Rough Pump Connect
Kit)
Isolation valve (included in the Alcatel pump kit)
External silencer (included in the Alcatel pump kit)
Noise reduction cover (included in the Alcatel pump kit)
To install the noise reduction cover
1.
Remove the clamp and blank flange from the pump inlet.
2.
Attach the DN40 nipple to the top of the inlet filter and o-ring assembly
and then secure the connection with a clamp.
3.
Attach the DN25 nipple to the outlet fitting and secure the connection
with a clamp.
4.
Connect the gray pump control cable to the rear of the pump.
Connecting the oil-free roughing pumpB-11